Please enable JavaScript to view this site.

GeoDict User Guide 2025

Filter Experiment Tab

Note-Info

Experimental setup and material parameters:

The oil passes through the pleated filter with a flow rate of 4 l/min. The filter has a surface area of about 280 cm², so the average superficial flow velocity in the filter media is therefore ~2.4 mm/s. Thus, flow in the filter media is slow.

The test reservoir volume is initially clean and contains 6 l. The injection reservoir volume contains 29 l which contain 11 g of the test dust. 25 ml are discharged per minute on the upstream and downstream side of the filter.

  1. In the Flow Settings panel on the bottom right, change the Flow Motion to Creeping flow (Stokes-Brinkman). Select as Inflow Material ID (ID 06) and Outflow Material IDs (ID 07).
  2. Select Multi Pass test – ISO 16889 Standard as Experiment Setup. In the diagram, adjust the pump flow rate to 4 l/min and leave the other parameters as the defaults shown below:
  3. In the Simulation Stopping Criterion panel. select Max. time reached, and set a time of 10 minutes.

©2026 created by Math2Market GmbH / Imprint / Privacy Policy